Fluke 8588A - Specific Function Calibration Verification

Fluke 8588A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reference Multimeter / 8½ Digit Multimeter
Calibration Verification
11
Digitize DC Voltage
Configuration: Z in DC Auto, Aperture = 200 ns, Trigger count 101010, Low pass filter Off.
Configuration: Z in DC Auto, Aperture = 200 ns, Trigger count 101010, Low pass filter 100 kHz.
Configuration: Z in DC Auto, Aperture = 200 ns, Trigger count 101010, Low pass filter 3 MHz.
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
1 V 0.000 000 0V 0.000 054 0 V 0.000 054 0 V 0.000 006 0 V
1 V 1.000 000 0V 0.000 157 0V 0.000 157 0V 0.000 006 1 V
1 V -1.000 000 0V 0.000 157 0V 0.000 157 0V 0.000 006 1 V
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
1 V 0.000 000 0V 0.000 019 0 V 0.000 019 0 V 0.000 006 0 V
1 V 1.000 000 0V 0.000 082 0V 0.000 082 0V 0.000 006 1 V
1 V -1.000 000 0V 0.000 082 0V 0.000 082 0V 0.000 006 1 V
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
1 V 0.000 000 0V 0.000 019 0 V 0.000 019 0 V 0.000 006 0 V
1 V 1.000 000 0V 0.000 082 0V 0.000 082 0V 0.000 006 1 V
1 V -1.000 000 0V 0.000 082 0V 0.000 082 0V 0.000 006 1 V