8845A/8846A
Calibration Manual
3-30
Table 3-18. 8845A/8846A Adjustment Steps (cont)
Step Modes Value Range Input Signal Description Series
48 ZIACS 1 10 A to Lo short AC 1 A Y
49 ZIAC 10 10 A to Lo short AC 10 A Y
AC ZIACS 10 10 A to Lo short AC 10 A N
50 Linearity
51 ACLIN 1.19 1.19 @1200 Hz AC 1 V N
52 ACLIN 0.8 0.8 @1200 Hz AC 1 V N
53 ACLIN 0.4 0.4 @1200 Hz AC 1 V N
54 ACLIN 0.005 0.005 @1200 Hz AC 1 V N
ACV
Gain
55 GVAC 0.1 0.1 @1200 Hz AC 100 mV Y
56 GVACS 0.1 0.1 @1200 Hz AC 100 mV N
57 ACPOLE 0.1 0.1 @50000 Hz A 100 mV N
58 GVAC 1 1 @1200 Hz AC 1 V Y
59 GVACS 1 1 @1000 Hz AC 1 V N
60 FVAC 1 1 @10 Hz AC 1 V N
61 ACPOLE 1 1 @50000 Hz AC 1 V N
62 GVAC 10 10 @1200 Hz AC 10 V Y
63 GVACS 10 10 @1200 Hz AC 10 V N
64 ACPOLE 10 10 @50000 Hz AC 10 V N
65 GVAC 100 100 @1200 Hz AC 100 V Y
66 GVACS 100 100 @1200 Hz AC 100 V N
67 ACPOLE 100 100 @50000 Hz AC 100 V N
68 GVAC 1000 1000 @1200 Hz
AC 1000 V (8845A use
750V on 750V range)
Y
69 GVACS 1000 1000 @1200 Hz
AC 1000 V (8845A use
750V on 750V range)
N
70 ACPOLE 1000 329 @50000 Hz
AC 1000 V (use 8845A
750V range)
N
VDC
Gain
71 GVDC 1000 1000 DC 1000 V N
72 GVDC -1000 -1000 DC 1000 V N
73 GVDC 100 100 DC 100 V N
74 GVDC -100 -100 DC 100 V N