8845A/8846A
Calibration Manual
A-2
Table A-1. Blank Verification Record for 90-Day Specifications
Nominal Output Test Limits
Ampl. Freq.
Range
High Low
Results Pass/Fail
8846A DC Volts
0 0.100 3.5E-6 -3.5E-6
100.0E-3 0.100 100.006E-3 99.994E-3
-100.0E-3 0.100 -99.994E-3 -100.006E-3
0 1 7.0E-6 -7.0E-6
1 1 1.000025 0.999975
-1 1 -0.999975 -1.000025
0 10 50.0E-6 -50.0E-6
5 10 5.000140 4.999860
-5 10 -4.999860 -5.000140
10 10 10.000230 9.999770
-10 10 -9.999770 -10.000230
0 100 600.0E-6 -600.0E-6
100 100 100.0033 99.9967
-100 100 -99.9967 -100.0033
0 1000 10.0E-3 -10.0E-3
1000 1000 1000.0410 999.9590
-1000 1000 -999.9590 -1000.0410
8845A DC Volts
0 0.100 3.5E-6 -3.5E-6
100.0E-3 0.100 100.0075E-3 99.9925E-3
-100.0E-3 0.100 -99.9925E-3 -100.0075E-3
0 1 7.0E-6 -7.0E-6
1 1 1.000037E+0 999.963E-3
-1 1 -999.963E-3 -1.000037E+0
0 10 50.0E-6 -50.0E-6
5 10 5.00015E+0 4.99985E+0
-5 10 -4.99985E+0 -5.00015E+0
10 10 10.00025E+0 9.99975E+0
-10 10 -9.99975E+0 -10.00025E+0
0 100 600.0E-6 -600.0E-6
100 100 100.0041E+0 99.9959E+0