917X Series Metrology Wells
Getting Started
2. To enable “Probe 1,” use the 4 or 5 button to turn the reference probe “On.” Press
“ENTER” to accept the selection.
Press the “Setup” soft key (Main Menu: R3. EF ME
NU|SETUP). In the reference probe
4. robe (ITS-90 or CVD) using the 4 or 5 keys. Press
5. ibration coefficients of the probe.
report of calibration that was shipped with
6.
u,
u: REF MENU|FOCUS). Use the 4 or 5 key to
n
setup menu, select “Probe 1” and press “ENTER”.
Select the calibratio
n type for the p
“ENTER” to accept the calibration type.
Use th
e numeric keys to enter the serial number and cal
Press “ENTER” after each parameter is entered to accept the ne
w parameter value.
Calibration coefficients can be found on a
your probe. If no coefficients can be found, contact th
e manufacturer or vendor of the
probe for assistance. If the probe is out of calibration, Fluke’s Hart Scientific Division
offers calibration services. Contact an Authorized Service Center for assistance on
obtaining a calibration for the probe.
After the reference probe has been enabled, the focus of the temperature display can be
set. The focus is disabled if the reference probe is disabled. From the reference men
press the “Focus” soft key (Main Men
select reference focus. Press “ENTER” to accept the selection. T
he largest temperature o
the display will now be the reference temperature.
18