Table 7. Sample Options
Option Description
Method
Time: The Product samples for the time recorded in this field.
Range: 00:00:01 to 23:59:59. In the Time-based sampling method, this value plus the hold time equals
one cycle.
Volume: The Product sample is equal to the volume recorded in this field. In the volume-based sample
procedure, this value, plus the hold time, is equal to one cycle. The sample stays at the estimated
value.
Hold The time between samples. Range: 00:00:00 to 23:59:59
Delay The time before the first sample starts after is pushed. Range: 00:00:03 to 23:59:59.
Cycles
The total of sample and hold intervals. Range: 0 to 999.
When in Automatic mode, the Product stops after the last cycle is complete. A value of 0 makes the Product
run until you push .
Mode
Automatic: the Product samples refer to the recorded parameters.
Manual: the Product samples once and stops.
Beep: Uses Automatic mode settings but ignores count alarm settings. The Product beeps for each particle
counted.