Manual Supplement BT520 Users
6 3/15
Set Measurement Thresholds
The Product lets you define upper and lower measurement thresholds or tolerance ranges. These
defined threshold values are then compared to the measured values to automatically identify and
prompt the user with a PASS, FAIL or WARN indicator of battery out of tolerance conditions.
The threshold function is disabled by default. You can set up to 10 set of thresholds and select one
threshold as needed.
To set and select measurement thresholds:
1. On the measurement screen, push the More softkey and Threshold softkey to open the Select
Threshold menu.
2. Use and to select one threshold set out of ten.
3. Use and L to highlight the value to be edited among Voltage lower, Reference, Warning,
Fail, Low limit, and Notes.
4. Edit the selected field.
a. Use the and + softkeys to change the values for Warning and Fail.
b. For other fields, push the Edit softkey, use the arrow keys to edit the value, and then push the
Confirm softkey to save the value.
5. Once all threshold values are correct, push the Confirm softkey on the save the threshold set.
The threshold set is applied and the T-X (X stands for value of No.) icon and the corresponding
PASS/WARN/FAIL indication shows on the display.
To disable measurement thresholds: