Data storage 18,000 data-points per channel
Data-collection start Start button, Temperature rise, Temperature fall, Date/time
Pre-trigger data stored Yes (configurable; see p. 28)
Hot-data protection By non-volatile memory, and software warning if reset attempted
before download
Communications USB
PC/software compatibility See p. 19
Logger reset Possible after download by start button (using previous run
parameters)
Battery Datapaq NiMH rechargeable battery-pack + lithium AA
Battery life Over 60 hrs continuous use (see p. 13), depending on sample
interval and operating temperature
Battery charger All regions except Americas – CH0070
Americas – CH0075
Specifications for Specific Thermocouple Types
Type J Type K Type N
Normal
range
Accuracy *
0°C to 800°C
32°F to 1,472°F
±0.3°C
±0.5°F
−100°C to 1,370°C
−180°F to 2,498°F
±0.3°C
±0.5°F
−100°C to 1,300°C
−180°F to 2,372°F
±0.4°C
±0.7°F
Lower range
Accuracy *
–
–
−190°C to −100°C
−310°F to −180°F
±0.5°C
±0.9°F
−190°C to −100°C
−310°F to −180°F
±0.5°C
±0.9°F
Resolution 0.1°C/0.2°F 0.1°C/0.2°F 0.1°C/0.2°F
Socket color
(IEC 584)
Black Green Pink
Type R Type S Type T
Normal
range
Accuracy *
50°C to 1,760°C
122°F to 3,200°F
±1.1°C
±2.0°F
50°C to 1,760°C
122°F to 3,200°F
±1.1°C
±2.0°F
−100°C to 400°C
−180°F to 752°F
±0.3°C
±0.5°F
Lower range
Accuracy *
−40°C to 50°C
−40°F to 122°F
±1.8°C
±3.2°F
−40°C to 50°C
−40°F to 122°F
±1.7°C
±3.1°F
−196°C to −100°C
−321°F to −180°F
±0.5°C
±0.9°F
Resolution 0.1°C/0.2°F 0.1°C/0.2°F 0.1°C/0.2°F
Socket color
(IEC 584)
Orange Orange Brown
*
Using sample interval > 0.8 s over temperature range shown. For more details, contact Datapaq.
DATAPAQ XDL12 Specifications and Operation 11