EasyManua.ls Logo

Fluke DATAPAQ XDL12 - Page 11

Fluke DATAPAQ XDL12
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Data storage 18,000 data-points per channel
Data-collection start Start button, Temperature rise, Temperature fall, Date/time
Pre-trigger data stored Yes (configurable; see p. 28)
Hot-data protection By non-volatile memory, and software warning if reset attempted
before download
Communications USB
PC/software compatibility See p. 19
Logger reset Possible after download by start button (using previous run
parameters)
Battery Datapaq NiMH rechargeable battery-pack + lithium AA
Battery life Over 60 hrs continuous use (see p. 13), depending on sample
interval and operating temperature
Battery charger All regions except Americas – CH0070
Americas – CH0075
Specifications for Specific Thermocouple Types
Type J Type K Type N
Normal
range
Accuracy *
0°C to 800°C
32°F to 1,472°F
±0.3°C
±0.5°F
−100°C to 1,370°C
−180°F to 2,498°F
±0.3°C
±0.5°F
−100°C to 1,300°C
−180°F to 2,372°F
±0.4°C
±0.7°F
Lower range
Accuracy *
−190°C to −100°C
−310°F to −180°F
±0.5°C
±0.9°F
−190°C to −100°C
−310°F to −180°F
±0.5°C
±0.9°F
Resolution 0.1°C/0.2°F 0.1°C/0.2°F 0.1°C/0.2°F
Socket color
(IEC 584)
Black Green Pink
Type R Type S Type T
Normal
range
Accuracy *
50°C to 1,760°C
122°F to 3,200°F
±1.1°C
±2.0°F
50°C to 1,760°C
122°F to 3,200°F
±1.1°C
±2.0°F
−100°C to 400°C
−180°F to 752°F
±0.3°C
±0.5°F
Lower range
Accuracy *
−40°C to 50°C
−40°F to 122°F
±1.8°C
±3.2°F
−40°C to 50°C
−40°F to 122°F
±1.7°C
±3.1°F
−196°C to −100°C
−321°F to −180°F
±0.5°C
±0.9°F
Resolution 0.1°C/0.2°F 0.1°C/0.2°F 0.1°C/0.2°F
Socket color
(IEC 584)
Orange Orange Brown
*
Using sample interval > 0.8 s over temperature range shown. For more details, contact Datapaq.
DATAPAQ XDL12 Specifications and Operation 11