©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 26
The expanded uncertainty
As mentioned, calculations for u
c
pertain to ± one standard
deviation of measurement uncertainties (covering 68% of the
population of measurements)
Usually it is desired to express uncertainty for a larger population or
condition, say 95% or 99%.
Expanding the calculated uncertainty through scaling estimates an
uncertainty that covers this larger population - U
m
.
A coverage factor, k, (often equal to 2), would indicate a 95%
confidence.
cku
U
m
68%
95%