©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 27
Now, returning to the
statement of uncertainty
... A measurement is complete only when
accompanied by a statement of the uncertainty of the
estimate. For example:
VDMM = 100.0051mV 0.0004 mV
In this case, 0.0004 mV would be the resulting value
of U
m,
calculated as shown below:
ckumV
U
m
0004.0
22
3
2
2
2
1
...
n
uuuuk