i3000/i2000 English Instruction Sheet
Page 4
Phase error < ±1° (45 Hz to 65 Hz),
±10° (at 20 kHz)
Linearity ±0.2 % of reading from 10 % to
100 % of range
Position sensitivity ±2 % of range with cable >25 mm
(1 inch) from the coupling
External field ±1 % of range with cable
>200 mm (8 in) from the head
Minimum load 100 k for specified accuracy
Noise
8 mV RMS (0.3 % of range) in
30 A range
2 mV RMS (0.1 % of range) in
300 A and 3000 A
Gain variation ±0.08 %/°C
Operating temperature
i3000s (24 and 36 Flex)
i2000 Flex
-20 to +85 °C (-4 °F to +185 °F)
0 to 70 °C (32 °F to +158 °F)
Storage temperature
i3000s (24 and 36 Flex)
i2000 Flex
-20 to +85 °C (-4 °F to +185 °F)
0 to 70 °C (32 °F to +158 °F)
Temperature coefficient: ± 0.08 % of reading / °C
Operating humidity: 15 % to 85 % (non condensing)
Altitude: 2000 m
Power supply Two AA IEC LR6 alkaline
batteries (all probes)
Battery life
i3000s (24 and 36 Flex)
i2000 Flex
400 hours typical
200 hours typical
Low battery Indicated by a red LED
General Characteristics
Output con.
i3000s (24 and 36 Flex)
i2000 Flex
Coaxial cable with a male
safety BNC connector. BNC to
Banana plug adaptor (supplied)
0.5 m cable with 4 mm safety
plugs
Weight 0.19 kg (0.4 lb)
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com