110/113/114/115/116/117
Calibration Information
6
Performance Tests
XW Warning
To prevent electric shock, do not perform the performance test procedures unless the Meter is
fully assembled.
The following performance tests verify the complete operation of the Meter and check the accuracy of each
Meter function against its specifications. The recommended calibration interval is 12 months. If the Meter fails
any part of the test, calibration adjustment and/or repair is indicated.
In the performance tests, the Meter is referred to as the device under test (DUT).
See Table 1 for a list of the features for each Meter.
Required Equipment
Table 2 is a list of the equipment required to conduct a performance test on the Meter.
Table 2. Equipment Requirements
Equipment
Measurement Function
Range
[1]
5522 Multi-product Calibrator (or equivalent)
DC Volts 10 mV to 600 V
DC Current 600 μA to 10 A
AC Volts 6 mV to 600 V
AC Current 600 μA to 10 A
Resistance 0 to 30 MΩ
Capacitance 9 to 900 μF
Temperature 0 °C to 400 °C
Frequency 2 V, 50 kHz
Fluke 80 AK
K-type Thermocouple Adapter Accessory
Temperature NA
K-type Thermocouple, mini-plug on both ends Temperature NA
Double Banana plug VoltAlert NA
[1] For accuracy specifications, see the Users Manual for the product.