EasyManua.ls Logo

Fluke Ti32 - Detailed Specifications

Fluke Ti32
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Thermal Imagers
Detailed Specifications
33
Power Saving...................................................... Sleep Mode activated after 5 minutes
of inactivity
Automatic Power Off after 30 minutes
of inactivity
Safety Standards
CAN/CSA............................................................ C22.2 No. 61010-1-04, UL STD
61010-1 (2
nd
Edition)
ISA ...................................................................... 82.02.01
Electromagnetic Compatibility ............................ Meets all applicable requirements in
EN61326-1:2006
Vibration ................................................................ 0.03 g
2
/Hz (3.8 grms), IEC 68-2-6
Shock ..................................................................... 25 g, IEC 68-2-29
Drop ....................................................................... 2 meter with standard lens
Size (H x W x L) ..................................................... 27.7 cm x 12.2 cm x 17.0 cm (10.9 in x
4.8 in x 6.7 in)
Weight.................................................................... 1.05 kg (2.3 lb)
Enclosure Rating .................................................. IP54
Warranty ................................................................ 2 years
Recommended Calibration Cycle........................ 2 years (with normal operation and
normal aging)
Supported Languages.......................................... Czech, English, Finnish, French,
German, Italian, Japanese, Korean,
Polish, Portuguese, Russian,
Simplified Chinese, Spanish, Swedish,
Traditional Chinese, and Turkish
Detailed Specifications
Temperature Measurements
Temperature Measurement Range (not calibrated below -10 °C)
Ti32, Ti29, Ti27............................................... -20 °C to +600 °C
TiR32, TiR29, TiR27 ....................................... -20 °C to +150 °C
Accuracy ............................................................. ±2 °C or 2 %, whichever is greater (at
25 °C nominal)
Measurement Modes .......................................... Smooth Auto-Scaling and Manual
Scaling
On-screen Emissivity Correction ........................ All models
On-screen Reflected Background
Temperature Compensation............................... All models
On-screen Transmission Correction .................. All models
Imaging Performance
Image Capture Frequency .................................. 9 Hz or 60 Hz refresh rate depending
on model