5. FORA 6 Total Cholesterol Test Strip
Accuracy
Results: Difference Plot Analysis
Number of subjects (N) = 320
TABLE 1 Ratio of samples (capillary whole blood) falling within different bias
ranges
Measurement range: 129 ~ 399 mg/dL
Linear Regression: y = 0.9777x + 1.9945, R² = 0.9724, (n=640)
User performance
Results: Difference Plot Analysis
Number of subjects (N) = 160
TABLE 2 Ratio of samples (capillary whole blood) falling within different bias
ranges
Measurement range: 129 ~ 399 mg/dL
Linear Regression: y = 0.9753x + 2.0318, R² = 0.9693, (n=320)
Precision
TABLE 3 Precision