Q33: Common Abnormal Results from Control Solution Test.
The results might be higher.
There is some residue on
the carrier (cap or tip) of the
control solution.
Clean the carrier (cap or
tip of control solution) and
test again. This time use
the second drop of the
control solution.
The results might be lower.
The pollutants are
absorbed in the control
solution.
Replace the control
solution and test again.
The results might be higher or
lower.
(1) It is not in the correct
user mode (e.g., not in
CTL or QC Mode
during the control
solution test).
(2) You may be using a
different control
solution.
(3) Control solution may
have expired.
(1) Choose the correct
user mode and test
again.
(2) Use ForaCare control
solution only.
(3) Discard 90 days after
opening.
The results might be lower.
Insufficient control solution
is absorbed.
Please test again and
remove the strip only after
hearing the "BEEP" sound
from the meter.
The variance might be larger.
The temperature of the
strips and meter is
unstable.
Please test again after the
strips and meter have
been placed in the same
environment for 30~60
minutes.