ForaCare FORA 6 Plus - Page 26

ForaCare FORA 6 Plus
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26/65
Tested sites
Difference within
±5mg/dL
Difference within
±10mg/dL
Difference within
±15mg/dL
Fingertip
26/45 (57.8%)
38/45 (84.4%)
45/45 (100%)
Palm
27/42 (64.3%)
41/42 (97.6%)
42/42 (100%)
Forearm
31/42 (73.8%)
39/42 (92.9%)
42/42 (100%)
Upper arm
29/42 (69.0%)
38/42 (90.5%)
41/42 (97.6%)
Table 2 Difference distribution for glucose concentration ≥ 100 mg/dL (5.55
mmol/L)
Tested sites
Difference within
±5mg/dL
Difference within
±10mg/dL
Difference within
±15mg/dL
Fingertip
61/115 (53.0%)
95/115 (53.0%)
115/115 (100%)
Palm
49/118 (41.5%)
92/118 (78.0%)
118/118 (100%)
Forearm
43/118(36.4%)
84/118 (71.2%)
115/118 (97.5%)
Upper arm
49/118 (41.5%)
87/118 (73.7%)
116/118 (98.3%)
Precision
In both intermediate precision and repeatability tests, the standard deviation (SD)
is within 5 mg/dL (0.28 mmol/L) for each glucose concentration < 100 mg/dL
(5.55 mmol/L) and the coefficient of variation (CV) is less than 5% for each
glucose concentration ≥ 100 mg/dL (5.55 mmol/L).