42
Important Control Solution Information
Use only ForaCare control solutions with your meter.
Do not use the control solution beyond the expiry date or 3
months after rst opening. Write the opening date on the control
solution vial and discard the remaining solution after 3 months.
It is recommended that the control solution test be done at
room temperature (68
°
F to 77
°
F / 20
°
C to 25
°
C). Make sure your
control solution, meter, and test strips are at this specied
temperature range before testing.
Shake the vial before use, discard the rst drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
Store the control solution tightly closed at temperatures between
35.6
°
F and 86
°
F (2
°
C and 30
°
C). Do NOT freeze.