7!
MENU OPTIONS
MENU: To enter the MENU mode, press the middle button when in the HOME
Detection Screen. The MENU screen is like the picture in the below left.
Shutdown: This is an alternative way to shutdown the device.
Setting: The setting looks like the image above right. There are several options in
the setting mode:
1. ChannelSet: Summarizes the sensor channel information for each
available sensor. It is also used to change/toggle the displayed
concentration unit of measure (i.e. ppm, mg/m
3
or other). Changes will
not be required as the most popular settings are always preset as default.
2. AirPump: This is a load calibration value. If a value >70, the pump is under
load and will beep to let the user know too much load is being experienced
by the pump. Check for any blockage or kinks in the tubing. One can also
SAVE the load value so the alarm warning is not activated.
3. ClearRec: Password 1111 enter. This option allows the user to clear the
alarm records in the memory of the device.
4. Language: Language options.
5. Backlight: Select the time for the screen to remain ON or Auto OFF delay.
6. SystemInf: Summarizes alarm points, and maximum detection limits.