6 TECHNICAL
SPECIFICATIONS
6.1 Safety
The insulation of external, inaccessible
circuits is reinforced.
6.2 Environmental conditions
The equipment is designed to be safe at least
under the following conditions:
·
Indoor use
·
Altitude up to 2000 m.
·
Temperature 5°C to 40°C.
·
Maximum relative humidity 80% for
temperatures up to 31°C decreasing linearly
to 50% relative humidity at 40°C.
·
Mains supply voltage fluctuations not
exeeding ±10% of the rated voltage.
·
Transient overvoltage is according to
category II, which is normal for this type
of equipment.
·
Pollution degree 2.
6.3 Lifting instructions
This instrument is not equipped with handles
or grips. One or two persons are needed to
lift the instrument, following the
instructions below:
1. Tilt the instrument slightly backwards,
lift and support the instrument as shown
in fig.12 to simplify and ensure a safe
lift.
2. Weight 23 kg.
1047 Soxtec System Hydrolyzing Unit, User Manual, Rev. 2.2, Part no 1000 3315, Foss Tecator AB, Sweden
13