12
Click the minutes (mm) and type 5 or use the arrows, and finally,
click the seconds (ss) and type 30 or select 30 using the arrows.
Displays the total recording time according to the logging interval
selected.
Select the Timer run check box if you want MicroLite to start
recording at a predetermined time. This option is convenient if
you are using several MicroLite at once and want them to all
begin logging at the same time.
Use the time and date selectors to set the start time.
In Cyclic run mode, the MicroLite overwrites the old
measurements (starting with the oldest recorded data) once the
MicroLite’s memory is full. Click the Cyclic run check box to
operate in this mode.
If the check box is unselected, MicroLite will operate in Normal
run mode and will stop recording when the memory is full.
In Push to run mode, the MicroLite will only start recording data
when you use the MicroLite magnet to close the reed relay on the
MicroLite’s circuit board, thereby causing the logger to start
recording data. This is convenient for when you wish to start
recording data soon (but not immediately) following setup, or at
an unspecified time.
2.3.2. Completing the Setup
Click Send Setup to send the new settings to the MicroLite and begin recording. This
will complete the setup. Click Cancel if you do not wish to change the setup at this
stage.
Note: The Send Setup command erases all existing data in the MicroLite.
If you select Timer run mode, MicroLite will wait in standby mode, displaying tRUN. It
will begin recording at the specified time.
If you select Push to Run mode, MicroLite will also wait in standby mode, displaying
PUSH until the magnet is used to start the logger.
2.3.3. Starting Data Recording
Click Run every time you want to start a new recording. The Run command erases
all previous data in the MicroLite memory and begins recording.
Note: If you setup the MicroLite using the Setup command, it will automatically begin
recording, and you don’t have to click Run.