What to do if a calibration error occurred on FujiFilm AMULET Innovality FDR MS-3500 Camera Accessories?
- Mmary84Aug 6, 2025
If a calibration error occurs with your FujiFilm Camera Accessories, conduct the daily calibration again.
What to do if a calibration error occurred on FujiFilm AMULET Innovality FDR MS-3500 Camera Accessories?
If a calibration error occurs with your FujiFilm Camera Accessories, conduct the daily calibration again.
Why images may not be obtained correctly on FujiFilm AMULET Innovality FDR MS-3500 Camera Accessories?
Images may not be obtained correctly with FujiFilm Camera Accessories due to a high temperature. Lower the room temperature to resolve this issue.
What to do if the door is open on FujiFilm Camera Accessories?
If the door is open on your FujiFilm Camera Accessories, close it before starting the exposure.
How to fix incorrect tube position in FujiFilm AMULET Innovality FDR MS-3500?
If the tube position is incorrect in your FujiFilm Camera Accessories, return the tube back to 0° before exposure.
What to do if the compression plate has not been set correctly on FujiFilm AMULET Innovality FDR MS-3500?
If the compression plate has not been set correctly on your FujiFilm Camera Accessories, reinstall the compression plate.
Brand | FujiFilm |
---|---|
Model | AMULET Innovality FDR MS-3500 |
Category | Camera Accessories |
Language | English |
Provides essential safety guidelines before operating the equipment.
Details safety precautions for using electric medical equipment correctly.
Covers general safety precautions, electrical shock, and radiation damage prevention.
Explains EMC standards and precautions for the exposure unit.
Outlines precautions for minimizing exposure and preventing accidents with the FDR MS-3500.
Shows the location of safety signs and labels on the exposure unit and diagnosis system.
Lists safety standard requirements for the diagnosis system.
Describes key features of the mammography diagnosis system.
Details the system's components, including exposure unit and AWS.
Identifies and explains the parts and functions of the exposure unit.
Identifies and explains the parts and functions of the diagnosis system.
Outlines fundamental procedures for starting up and operating the system.
Provides step-by-step instructions for removing and attaching compression plates.
Explains how to use emergency stop buttons to halt equipment operation safely.
Describes how the device enters and exits sleep mode to reduce load.
Offers solutions to common problems based on AWS screen messages.
Displays common system messages and their remedies for quick problem resolution.
Covers daily checks for the exposure unit, including inspection before and during studies.
Outlines weekly and monthly inspection procedures for system components.
Lists periodical maintenance and replacement parts handled by authorized dealers.
Details technical specifications for the exposure unit, including power and operating characteristics.
Describes radiation field sizes and how to check them with different compression plates.
Explains the coverage, modes, and confirmation of the Automatic Exposure Control (AEC).
Details the procedure for measuring Average Glandular Dose (AGD) using an acrylic-board phantom.
Shows storage environment limitations for system components.
Presents physical testing data including sensitometric response, spatial resolution, and noise analysis.
Summarizes the study objective, design, and methodology for clinical image evaluation.
Explains the purpose, preparation, installation, and confirmation of the magnification table.
Provides instructions for attaching the chest wall pad to the exposure table.
Provides instructions for attaching the axilla pads to the exposure table.
Describes how to connect the remote exposure switch to the control pad.
Explains how to attach and detach the shielding plate for radiation protection.
Details how to attach and detach the Face Guard T.