Installation and configuration
2001989-351A ApexPro™ 3-21
7. Proceed to the next digit by pressing the Graph button until the next LED
flashes.
8. Repeat for all 4 (5 for T14 transmitter) digits.
The TTX number is not changed until the fourth/fifth digit is accepted by
pressing the Graph button. The top row LEDs flash twice upon acceptance of the
TTX number.
9. To exit service mode, remove the short from pins 2 and 4. The service program
runs until the short is removed. Then the transmitter reloads the application code
and continues to run normally at the new TTX number.