7-6 ApexPro™ 2001989-351A
Checkout
NOTE
Type help at any time to see a list of commands.
2. Verify new prompt, Unity MC IP address of the ApexPro application, then
C:Program Files\Marquette\PTS\X.X.
3. Type display receiver and press Enter.
4. The number of beds correlates to the number of receivers indicated by the
number of green Receiver Slots LEDs on the back of the Receiver System. Four
receivers per one green Receiver Slots LED.
Test available receivers
1. Type admit <space> care unit name <space> bed name <space> patient
name <space> TTX# <space> patient age, where
care unit name = a temporary unique name for testing
bed name = a temporary unique name for testing
patient name = test
TTX# = number of transmitter using for test (Use the number in parentheses on
the transmitter label or in the CIC drop-down menu, e.g., 1071.)
age = 55
pstconfig2
1 LED = 4 receivers
2 LEDs = 8 receivers
3 LEDs = 12 receivers
4 LEDs = 16 receivers