2001989-351A ApexPro™ 7-5
Open prompt
1. Connect a patient simulator to a transmitter.
2. Depending on where the ApexPro application resides, open a command prompt.
See ApexPro application residing on a CIC on page 5-15. or See ApexPro
application residing on an ATS on page 5-16..
3. In the MS-DOS command window type
cd <space> C:\Program Files\Marquette\PTS\X.X> (where X.X is the current
ApexPro software version)
4. Verify the prompt
C:\Program Files\Marquette\PTS\X.X>
Determine number of admitted beds
1. At the prompt type unityviewer.
The Unity Viewer window opens.
The message Waiting for network traffic... appears.
2. In the displayed information, identify the TELE TOWER with the arrow
preceding the IP address and click on the + (on the left) to expand.
NOTE
The small arrow preceding the IP address identifies the tele tower (i.e. the
ApexPro Telemetry System) currently being checked for connectivity. This
arrow is highlighted in the following graphic.
unityviewer
3. This displays the number of beds admitted on the ApexPro Telemetry System.
Note this number and close the window.
Determine number of unassigned receivers
1. In the MS-DOS window at the PTS\X.X prompt, type ptsconfig and press Enter.

Table of Contents

Other manuals for GE ApexPro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ApexPro and is the answer not in the manual?

GE ApexPro Specifications

General IconGeneral
BrandGE
ModelApexPro
CategoryMedical Equipment
LanguageEnglish

Related product manuals