4-14 ApexPro™ 2001989-351A
Maintenance
504A
d. Click OK.
5. Identify and record the band that the transmitter is operating from the following
table:
Band: ___
6. Type the command
:34<space>X<space>1
NOTE
X is the frequency band recorded in step 5 (e.g. :34 1 1 for band 1)
7. Record the initial pot value shown:
______________
T14 frequency bands TTX range
Band 0 1395.025-1395.975 10199-10161
Band 1 1396 - 1396.975 10160-10121
Band 2 1397 - 1397.975 10120-10081
Band 3 1398 - 1398.975 10080-10041
Band 4 1399 - 1399.975 10040-10001