4-16 ApexPro™ 2001989-351A
Maintenance
4. Configure the ASCII setup.
a. Select File > Properties.
b. In the Settings tab, click ASCII Setup...
c. Verify the settings are the same as shown:
401A
d. Click OK.
5. Press Enter.
6. When prompted, type the password mms_aps and press Enter.
7. At the @> prompt, type ss and press Enter.
8. Determine if patients are admitted to the receiver, use the following graphics for
reference:
If the column next to FQ is all -1 values, no beds are admitted on the receiver
subsystem tested. Proceed to step 9.
Baud-Rate 19200
Data Bits 8
Parity None
Stop Bit 1
Flow Control Xon Xoff