Maintenance
2001989-351A ApexPro™ 4-17
402A
If the column next to FQ is displays any hertz values, each hertz value
represents an admitted bed on the receiver subsystem tested. Proceed to step
10.
403A
9. If no beds are admitted, perform the following steps:
a. Type l<space>1<space>1 and press Enter.
NOTE
The command is lower case letter L <space> number one <space>
number one. This will load application one to receiver one.
b. Type sf<space>1<space>600000000 to admit a reserved frequency to
receiver one.
NOTE
The command is sf<space>number one<space> number six followed by
eight zeros.