7-1
Chapter 7
Specifications
Note: All specifications are nominal and are subject to change without notice.
Electrical Specifications
Device Insulation Class
Class II Double Insulated
Applied Parts (Light Pad)
Type BF
Input Voltage
100 240 V~, 50/60 Hz
Input Current
0.25A 0.75A
Leakage Touch Current
< 100 μA in normal condition
< 500 μA in single fault condition
Environmental Operating Conditions
Temperature
+5°C to +35°C (41°F to 95°F).
Humidity
15% to 90% RH, non-condensing, but not requiring a water
vapor partial pressure > 5 kPa
Atmospheric pressure
70 kPa to 106 kPa (10,000 ft. to -1,250 ft.)
Light Box
IP21
Light Pad
IPX4
Transportation Conditions & Storage Conditions (while sealed in shipping carton)
Temperature
-25°C to +5°C (-13°F to 41°F) without humidity control.
Temperature
+5°C to +35°C (41°F to 95°F) up to 90% RH,
non-condensing
Temperature
+35°C to +70°C (95°F to 158°F) at a water vapor
pressure up to 5 kPa