Interfacing with Non-GE Devices
2000966-386D Dashâ„¢ 3000/4000/5000 E-3
WARNING
SINGLE PATIENT USE—All eight serial ports on the connectivity
device must only be used by ONE patient.
CAUTION
INSTALLATION—To avoid accidental ingress of liquids, always
mount the connectivity device in a vertical position with the
connectors at the bottom.
CAUTION
TREATMENT—Do not treat a patient based solely on the alarm
messages and/or numerics presented via the connectivity device to
the monitor. You must verify the accuracy of the alarm message and/
or numerics at the peripheral device itself before initiating treatment.
Treatment should be based on the information presented at the
peripheral device.
Non-GE Device Information
Displayed Data
The peripheral device data that displays on the monitor varies by device, as shown in
the following table.
The monitor only displays the information the peripheral device sends to the
connectivity device. The parameters sent vary with each external device and are
subject to change without notice by the manufacturer.
Device Waveform
Parameter
Window
Trends
Alarm
Broadcast
Graphing
Anesthesia machines Yes Yes Yes Yes Yes
Continuous cardiac output No Yes Yes Yes Yes
Gas analyzers Yes Yes Yes Yes Yes
IV pumps No No No Yes No
Pulse oximeters No Yes Yes Yes Yes
Transcutaneous monitors No Yes Yes Yes Yes
Urometers No Yes Yes No Yes
Ventilators Yes Yes Yes Yes Yes
BIS NoYesYesNoYes