4400-0024 enFlow Operator’s Manual EN Rev. T 05/12  Page 21 of 26
     
 
Guidance and Manufacturer’s Declaration - Emissions 
The enFlow 100 with enFlow 120 is intended for use in the electromagnetic environment specified below. The 
customer or user of the enFlow 100 with enFlow 120 should ensure that it is used in such an environment. 
Electromagnetic Environment – Guidance 
RF Emissions 
Radiated 
CISPR 11 
The enFlow 100 with enFlow 120 uses RF energy only for its internal 
function. Therefore, its RF emissions are very low and are not likely to 
cause any interference in nearby electronic equipment. 
The enFlow 100 with enFlow 120 is suitable for use in all establishments, 
other than domestic, and those directly connected to the public low 
voltage power supply network that supplies buildings used for domestic 
purposes. 
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity 
The enFlow 100 with enFlow 120 is intended for use in the electromagnetic environment specified below. The 
customer or user of the enFlow 100 with enFlow 120 should ensure that it is used in such an environment. 
Electromagnetic Environment – Guidance 
Level 3 (±6kV) 
Contact 
 
Level 3 (±8kV) 
Air 
Level 3 (±6kV) 
Contact 
 
Level 3 (±8kV) 
Air 
Floors should be wood, concrete or 
ceramic tile. If floors are synthetic, the 
r/h should be at least 30% 
Mains power quality should be that of a 
typical commercial or hospital environment. 
±1kV Differential 
±2kV Common 
 ±1kV Differential 
±2kV Common 
Mains power quality should be that of a 
typical commercial or hospital environment. 
Voltage 
Dips/Dropout 
EN/IEC 61000-4-11 
>95% Dip for 
0.5 Cycle 
 
60% Dip for 
5 Cycles 
 
30% Dip for 
25 Cycles 
 
>95% Dip for 
5 Seconds 
100% Dip for 
0.5 Cycle 
 
60% Dip for 
5 Cycles 
 
30% Dip for 
25 Cycles 
 
See Note 1 
Mains power quality should be that of a 
typical commercial or hospital 
environment. If the user of the enFlow 
100 with enFlow 120 requires 
continued operation during power mains 
interruptions, it is recommended that 
the enFlow 100 with enFlow 120 be 
powered from an uninterruptible 
power supply or battery. 
Power Frequency 
50/60Hz 
Magnetic Field 
EN/IEC 61000-4-8 
Power frequency magnetic fields 
should be that of a typical commercial 
or hospital environment. 
Note 1) During the 5 Second event, the enFlow 100 and 120 power off, but return to normal operation as soon as 
power is restored. 
Guidance and Manufacturer’s Declaration – Emissions 
The enFlow 100 with enFlow 120 is intended for use in the electromagnetic environment 
specified below. The customer or user of the enFlow 100 with enFlow 120 should ensure that 
it is used in such an environment. 
Electromagnetic Environment – Guidance 
 
Conducted RF 
EN/IEC 61000-
4-6 
 
Radiated RF 
EN/IEC 61000-
 
3 Vrms 
150 kHz to 80 MHz 
 
3 V/m 
80 MHz to 2.5 GHz 
 
Portable and mobile communications 
equipment should be separated from 
the enFlow 100 with enFlow 120 by no 
less than the distances calculated/listed 
below: 
 
D=(3.5/3V/m)(Sqrt P)