4400-0024 enFlow Operator’s Manual EN Rev. T 05/12  Page 22 of 26
     
 
D=(3.5/3V/m)(Sqrt P) 
80 to 800 MHz 
 
D=(7/3V/m)(Sqrt P) 
800 MHz to 2.5 GHz 
 
where P is the max power in watts and 
D is the recommended separation distance 
in meters. 
 
Field strengths from fixed transmitters, 
as determined by an electromagnetic 
site survey, should be less than the 
compliance levels (3Vrms and 3V/m). 
Interference may occur in the vicinity of 
equipment containing a transmitter. 
Recommended Separation Distances Between Portable and Mobile RF Communications 
Equipment and the enFlow 
The enFlow 100 with enFlow 120 is intended for use in the electromagnetic environment in which radiated 
disturbances are controlled. The customer or user of the enFlow 100 with enFlow 120 can help prevent 
electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications 
Equipment and the enFlow 100 with enFlow 120 as recommended below, according to the maximum output power 
of the communications equipment. 
Separation (m) 
150kHz to 80MHz 
D=(3.5/V1)(Sqrt P) 
Separation (m) 
80 to 800MHz 
D=(3.5/E1)(Sqrt P) 
Separation (m) 800MHz to 2.5GHz 
D=(7/E1)(Sqrt P)