8 Technical specifications
Regulation Maximum voltage and current with automatic cross-over
at preset limits
Output mode Contante Voltage: 0–300 V DC,
Contante Current: 0–400 mA
Maximum Power Output 80 W
Programming range Voltage: 5–300 V DC
Current: 10–400 mA
Time: 00.01–23.59 h or off (- - - -)
Output resolution Voltage: 1 V
Current: 1 mA
Programming resolution Voltage: 1 V
Current: 1 mA
Time: 1 min
Accuracy Voltage: 4%, ± 2 V
Current: 4%, ± 4 mA
Timer: 0.1% ± 1 min
Line regulation < 0.5%
Load regulation < 1% at load change 10–90% of maximum load
Ripple < 3% at 300 V
Short term stability < 0.2% /10 h after warm up
Long term stability < 1% /year
Output protection Fully protected against any overload conditions
Recovery after power failure The program continues automatically
Ambient operating temperature 4–40 °C
Ambient operating humidity 0–95%
Ambient operating pressure 68–106 kPa, maximum altitude of 2000 m
Mains requirements 100–120 V/220–240 V; 50/60 Hz
Power consumption Max 120 W
Dimensions (WxDxH) 250 x 215 x 95 mm
Weight 3.0 kg