4 Operation
4.3 Setting up a run
30 Electrophoresis Power Supply EPS 3501 XL Operating Instructions 28-9639-69 AA
Table 4-1 and Figure 4-2 illustrate a step program. The programming and running of this
application are shown as Figure 4-1 and Figure 4-5.
Figure 4-2. Programming the voltage limiting profile in STEP mode. The parameters shown are the
same as those listed in Table 4-1.
Table 4-1. The parameters of a step program.
Choosing gradient ( ), means that a voltage endpoint for the actual phase is
programmed together with current and power limiting values. A linear voltage
gradient is made with zero (for the first phase) or the programmed endpoint of the
phase before (for the next phases) as starting point and the programmed endpoint as
No. Description No. Description
A
Phase 1
D
Phase 4
B
Phase 2
E
Voltage (V)
C
Phase 3
F
Time (h)
Phase number Voltage (V) Current (mA) Power (W) Volthours (Vh)
1100010101000
2500105120
3200020151000
4350030251500
500
1000
2000
3500
1000 Vh
120 Vh
1000 Vh
1500 Vh
12
E
F
A
B
C
D