Electrophoresis Power Supply EPS 3501 XL Operating Instructions 28-9639-69 AA 37
Operation 4
Performing a run 4.5
Pausing
You can interrupt the electrophoresis for sample loading and/or changing the program
by pressing pause continue. Voltage will no longer be supplied, the HV on LED goes off,
and you may safely load your samples. The display shows the status of the run when
pause continue was pressed.
Switch between time, integrated voltage and integrated current for the phase running
when pause continue was pressed with .
When sample loading is complete, press either pause continue again or run to
continue the run from where it was interrupted.
Editing a running program when in PAUSE
When in the PAUSE mode you can also press set enter to make changes in the program.
This mode is called the P-SET mode. When the P-SET mode is entered you can make
changes as described in Changing a parameter value‚ on page 33. It is not possible to
insert or delete a phase for a running program. The P-SET mode is the same as the SET
mode apart from restrictions in setting the breakpoint. Naturally, it is not possible to
enter a time, integrated voltage or current that is already passed.
Press exit or pause continue to go back from P-SET to PAUSE. Press run or pause
continue to proceed with the electrophoresis.