5 Operation
16
Electrophoresis Power Supply EPS 601 User Manual 18-1130-19 Edition AD
SET: P1 OFF
200 V 40 mA 10 W
enter
Select by
Confirm by
SET: P2 OFF
6 V 400 mA 100 W
enter
Select by
Confirm by
SET: P2 OFF
600 V 400 mA 100 W
enter
Select by
Confirm by
SET: P2 OFF
600 V 50 mA 100 W
enter
Select by
Confirm by
SET: P2 OFF
600 V 50 mA 30 W
enter
Select by
Confirm by
SET: P2 0:00 h
600 V 50 mA 30 W
enter
Select by
Confirm by
Alarm at End ? Y
enter
Select by
Confirm by
SET: P2
A
2:00 h
600 V 50 mA 30 W
SET: P2 OFF
600 V 50 mA 30 W
1. Choose Program
number
2. Set Voltage limit
3. Set Current limit
4. Set Power limit
5. Choose Breakpoint
unit (h or Vh) or no
Breakpoint (OFF).
6. Set Breakpoint
7. Choose Alarm
YES or NO
8. Program ready
Breakpoint off
Fig. 3.Step-by-step summary of programming.