Technical Specifications 8
8 Technical Specifications
Regulation Maximum voltage, current and power with automatic cross-
over at preset limits
Output range Voltage: 0–600 V DC
Current: 0–400 mA
Power: 0–100 W
Programming range Voltage: 6–600 V DC
Current: 1–400 mA
Power: 1–100 W
Time: 00:01–500 h
Volthour: 1–500 000 Vh
Output resolution Voltage: 1 V
Current: 100 μA, 0–39.9 mA
1 mA, 40–400 mA
Power: 1W
Programming resolution Voltage: 1 V
Current: 1 mA
Power: 1 W
Time: 1 min, 00:01–99:59,
1h, 100–500 h
Volthour: 1 Vh, 1–9999 Vh,
100 Vh, 10.0–99.9 kVh
1 kVh, 100–500 kV
Accuracy Voltage: 1%, ±3 V
Current: 1%, ±1 mA
Power: 2%, ±1 W
Timer: 0.1% ±1 min, 00:01–99:59 h
0.1%, ±10 min, 100–500 h
Line regulation < 0.2%
Load regulation < 1% at load change 10–90% of maximum load
Ripple < 1% at 600 V
Short term stability < 0.2% /10 h after warm up
Long term stability < 1% /year
Start current check Resistance not greater than 0.2 Mat 40 V
(current less than 200 μA). Can be disabled
Sudden load change check
Ground leakage check Leakage not greater than 500 μA
Output protection Fully protected against any overload conditions
Recovery after power failure Duration < 8 s, The program continues automatically
Duration ≥ 8 s, The program continues after manual restart
27
Electrophoresis Power Supply EPS 601 User Manual 18-1130-19 Edition AD