Control Handle
OFF
■ Shuts off water.
FILTER
■ For filtered water. Directs water flow
through filter.
BYPASS
■ Permits flow of water to the house without flowing through the filter.
Use this position for filter changes or if filtered water is not desired
.
2
Filter Cartridge Replacement
You should change your filter when the water flow is noticeably reduced or at least every 3 months.
Turn off water to filter. Rotate the control handle to the
BYPASS or OFF position (see
Valve Operations
section).
Press the red pressure-release button to release pressure.
Unscrew the filter canister and discard used filter. Wash the
filter canister with mild soap and water.
Do not
use harsh
cleaners or hot water.
Inspect the filter canister O-ring. Make sure it is lightly
lubricated with clean food-grade silicone grease (silicone
grease is available through GE Parts and Service:
1.800.626.2002, part number WS60S10005). Be sure the
O-ring is seated in the groove. It is recommended that you
replace the O-ring if it is damaged.
Place a new filter cartridge into the canister, making sure it
is centered and completely seated on the bottom seal.
Reinstall the filter canister to the unit. Use the canister
wrench to tighten the canister. DO NOT OVERTIGHTEN.
Slowly rotate the control handle to the FILTER position.
Press the red pressure-release button to remove trapped air.
After installation, open any downstream faucet and flush the
cartridge for 15 minutes, wait one hour, then flush again for
15 minutes before using the water.
9
8
7
6
5
4
3
2
1
Installation Overview
NOTE: Be sure to allow a minimum space
of 1
1
⁄
2
″
–2
″
under the filter for removing
the sump to change the cartridge.
Turn canister wrench
clockwise to remove
Canister
Turn
clockwise
to remove
canister
Turn
counterclockwise
to tighten
Red pressure-
release button
Filter
cartridge
O-ring seal
Filter canister
Head
STEP-BY-STEP INSTALLATION INSTRUCTIONS
Recommended
shut-off valve
Recommended
shut-off valve
Fitting
Ferrule
Hex nut
Fitting
Ferrule
Hex nut
1
1
⁄2 ″–2″
Typical installation
Remote timer
Mounting
bracket
CARTRIDGE SPECIFICATIONS
Influent Challenge Reduction Average
Substance Concentration Requirements Reduction
Standard 42
Chlorine 2.0 mg/L ± 10% ≥ 50% 87.8%
Performance Data
This system has been tested according to NSF/ANSI 42
for the reduction of the substances listed at right. The
concentration of the indicated substances in water
entering the system was reduced to a concentration
less than or equal to the permissible limit for water
leaving the system, as specified in NSF/ANSI 42.
FXHTC Cartridge
Test Conditions
Flow Rate: 3 gpm (11.4 Lpm)
pH: 7.5± 1
Inlet Pressure: 60 PSI (4.1 bar)
Temperature: 68°F ± 5°F (19.8°C ± 2.5°C)
Operating Requirements
Pressure: 30–100 PSI (2.1–6.9 bar)
Turbidity: 5 NTU Max.
Temperature: 40°F–100°F (4.4°C–37.7°C)
Capacity: 30,000 gallons (113,562L) or 3 months
Testing was performed under standard laboratory conditions; actual performance may vary.
Control handle