EasyManuals Logo

GE LightSpeed 16 User Manual

GE LightSpeed 16
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
g
Product Data Sheet – rev 4.2
July 4th 2003
Page 14
LightSpeed
16
With Xtream
CT Scanner System
GE Medical Systems-America
Milwaukee, USA - Fax: 1 414 544 3384
GE Medical Systems-Asia:
Tokyo, Japan - Fax: 81 425 85 5490
Hong Kong - Fax: 852 2559 3588
GE Medical Systems-Europe
Peripherals:
Heat Storage Capacity: 6.3 MHU 16 x 1.25 mm (uses all 24 rows)
16 x 0.625 mm (uses center 16 rows )
Heat Dissipation:
Total of 254 GB system:
Anode (max) 840 KHU/min
70% geometric efficiency; 98% absorption
efficiency.
Casing (cont) 300 KHU/min
Main system (host) disk drive:
Tube Unit: 6.9 kW continuous for 10
minutes
High Performance Drive
Data Acquisition System
36 GB, 3.5 inch form factor
Dual Focal Spots:
15,000 RPM
Small Focal Spot:
12,288 available input channels.
Ultra320 SCSI interface
0.7 mm (W) x 0.6 mm (L) nominal
value
(IEC 336/93)
Assigned to applications software and
image files
1,640 Hz maximum sample rate.
Effective analog to digital conversion range
greater than two million to one.
2 system disk drives (Image Disk)
0.9 mm (W) x 0.7 mm (L) traditional
methodology
High Performance Drive
73 GB, 3.5 inch form factor each.
Scan/Control Unit
15,000 RPM
Large Focal Spot:
0.9 mm (W) x 0.9 mm (L) nominal
value
(IEC 336/93)
Ultra320 SCSI interface
Located in base of Operator Console.
Assigned to image files only
- 250,000 uncompressed 512 images
1.2 mm (W) x 1.2 mm (L)
traditional methodology
Scan data disk drive:
Host Computer
High Performance Drive
36 GB, 3.5 inch form factor
Maximum Power: 53.2 kW
Dual SMP 2.66.GHz Intel Xeon processors
with 512KB L2 cache.
UltraSCSI interface
Beam collimated to 55° fan angle.
Intel Hyper-threading technology.
Scan data disk drive:
Average time to replace tube: <
10 hours
High Performance Drive
2GB DDR266 Dual Channel Memory with a
throughput of 4.2GB/sec
36 GB, 3.5 inch form factor
High Voltage Generation
UltraSCSI interface
Image Processor:
High-frequency on-board generator.
Continuous operation during scan.
Standard MOD drive:
Nvidia Quadro4 980XGL AGP 8X graphics
with 128MB Memory
Magnetic Optical Disk Drive
53.2 kW output power.
Erasable, rewritable media
kVp: 80, 100, 120, 140
Graphics Processor Unit (GPU) Clock
300Mhz
2.3 GB, 3.5 inch form factor
mA: 10 to 440 mA, 5 mA increments
Assigned to DICOM 3.0 format image
file.
Graphics Memory Clock 325Mhz
Maximum mA for each kVp selection:
Stores 4,700 lossless JPEG compressed
512 image files per side
Image Reconstruction Engine
(GRE)
KVp Max mA
80 400
100 420
120 440
140 380
Off-line retrieval of image. Images may
be viewed as soon as they are restored
from MOD
Custom-designed special purpose
CT Image Generator
DVD Ram:
Custom CT back projection hardware
accelerates 2D & 3D back projection.
4.7 GB per side, 5.25" half height form
factor
HiLight Matrix II Detector:
Intel Hyper-threading Technology.
Transfer rate 2.7MB/sec
32-bit floating point data format
Assigned to scan data file and protocol
file storage/retrieval.
21,888 individual elements composed by: 8
rows of 1.25 mm thickness and 16 rows of
0.625 mm thickness, each containing 888
active patient elements; 24 reference
elements.
2GB DDR226 ECC Dual Channel
Memory Standard (4.2 GB/sec).
Color monitors:
Software Architecture:
21 inch diagonal width
4 typical modes of data output:
1280 x 1024 dot resolution
Software architecture based on industry
standards and client-server design
8 x 1.25 mm (uses center 16 rows)
Non-interlaced, flicker-free presentation
8 x 2.5 mm (uses all 24 rows)
76 kHz Horizontal deflection frequency
* Option

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LightSpeed 16 and is the answer not in the manual?

GE LightSpeed 16 Specifications

General IconGeneral
TypeCT Scanner
Number of Slices16
Rotation Time0.5 seconds
Gantry Aperture70 cm
Detector TypeSolid-state detector
Scan TypeHelical and Axial
X-ray Tube Voltage80-140 kV
X-ray Tube Current10-440 mA
Display Matrix512 x 512
Power Requirements480 VAC, 3 Phase, 60 Hz, 100 kVA