g 
Product Data Sheet – rev 4.2 
July 4th 2003  
Page 5 
 
 
 
LightSpeed
  
With Xtream 
CT Scanner System 
GE Medical Systems-America 
Milwaukee, USA  - Fax: 1 414 544 3384 
GE Medical Systems-Asia: 
Tokyo, Japan - Fax: 81 425 85 5490 
Hong Kong - Fax: 852 2559 3588 
GE Medical Systems-Europe 
Scan Fields of View
: 
Biopsy:  Simplified prescription for single 
or multiple scans around an arbitrary table 
position aids biopsy studies.  
 
—  25 cm for adult head 
16-SLICE HELICAL MODES 
Table Speed (mm/rotation) 
Slice 
Thick-
ness 
(mm) 
Pitch 
0.5625  
:1 
 
 
0.9375 
:1 
 
 
1.375
:1 
 
 
1. 75:1 
 
0.625 5.625 9.375 13.75  17.5 
1.25   5.625 
 11.25 
9.375 
18.75  
13.75  
27.5 
17.5  
35 
2.5   5.625 
 11.25 
 9.375 
18.75  
13.75  
27.5 
17.5 
35  
3.75    5.625 
 11.25 
9.375 
18.75 
13.75  
27.5 
17.5 
35 
5 5.625 
 11.25 
9.375 
18.75 
13.75  
27.5 
17.5 
35 
7.5 11.25 18.75 27.5  35 
10 11.25 18.75 27.5  35 
—  25, 50 cm for body 
Advanced Artifact Reduction (AAR) 
Filter significantly reduces streaking 
artifacts when highly absorbent objects 
are in the field of view – ie: large 
shoulders, screws. 
—  25 cm for pediatric head 
Helical Scan Enhancements 
Helical Image Reconstruction 
Full simultaneity allows complete image 
display, processing and analysis, as well 
as image archival and filming, concurrent 
with scanning and reconstruction -- even 
when acquiring helical images in a multi-
slice mode. 
Reconstruction Algorithms: Soft Tissue, 
Standard, Detail, Bone, Bone Plus, Lung 
and Edge 
Reconstruction Matrix
: 512 
Confirm Rx to X-Rays on: < 15 sec. for 
any state of tube and gantry; < 10 sec. 
with the gantry rotating 
Display Matrix
: 1024. 
Display FOV
: Freely variable center/off-
center, prospective/retrospective target 
selection. 
Anatomical Programmer: a ten-region 
anatomical selector allows quick and easy 
access to 90 user-programmable 
protocols per region.  Separate selector 
for adult and pediatric exams.  There are 
four selection tabs to select: GE, User, 
Service and Most Recent Patient.  
Copy/Paste is supported for easy 
modification and copying of protocols. 
CT Number Scale
: -1024 to 3071 HU 
Helical Reconstruction Times: 
 
  Reconstruction time as fast 6 
images per second 
Helical Scan Parameters 
  Typical 0.167 sec. image-to-image 
recon in normal 16 slice recon mode. 
Ten user-defined regions.  Each region 
has one default protocol displayed with the 
anatomical selector for very fast access to 
most commonly used protocols 
Scan Speeds: Full 360° rotational scans in 
0.5, 0.6, 0.7, 0.8, 0.9, 1.0 seconds.  
  Maximum image-to-image cycle 
time is ± 10% for prospective and 
retrospective image-to-image display. 
This applies for 512 matrix; any 
display FOV; in AutoView (all 
layouts); with concurrent filming and 
image archival for all scan modes. 
Scan Technique
: 
Protocols include preset scan time, kVp, 
mA, scan mode, slice thickness and 
spacing, table speed, scan FOV, display 
FOV and center, recon algorithm and 
special image acquisition and processing 
options   
—  kVp:   80, 100, 120, 140 
—  mA: 10 to 440 in 5 mA increments 
—  Power:  0.8 to 53.2 kW 
  Iterative bone processing 
increases time by 250 milliseconds. 
Iterative bone processing, which is 
always enabled for adult head 
scanning, reduces image artifacts in 
head scans stemming from X-ray 
beam hardening effects. 
—  Focal Spot Selection: 
Any scan parameters may be edited for 
each scan or all scans - either before or 
during an exam.  The number of scans 
may also be easily changed. 
    Small spot for up to 24 kW 
    Large spot for greater than 24 kW 
Single Acquisition
: 120 second scan 
maximum. 
AutoScan™: Fully automates longitudinal 
table movement and start of each scan. 
Minimum DFOV:
 9.6 cm 
Helical Tilt
: helical acquisition is possible 
with the gantry tilted to a maximum of 30 
degree, in half degree increments. 
AutoVoice™: 3 preset (English) and 17 
user-defined messages automatically 
deliver patient breathing instructions; 
especially useful for multiple helical 
scanning. 
Minimum Pixel Size: 
0.19 mm 
Queued Recon
: Requests will be 
processed continuously and 
simultaneously with other processes on 
the system including scanning.  
Prospective recon will be prioritized over 
retrospective recon. 
Multiple Acquisition Maximum
 Scan Time: 
Multiple scans may be acquired in one 
series to produce up to 3,000 contiguous 
helical images. Up to 2,881 seconds 
helical coverage is possible in multiple 
series. 
Trauma Patient: Allows patient scans and 
image display/analysis without entering 
patient data before scanning. 
Priority Recon Queuing
: One touch 
selection marks most recent rotation for 
Minimum Inter-Group Delay
 (IGD): 5 
seconds between adjacent helical scans 
* Option