GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
8-6 Section 8-3 - Loading the System Software
8-3-1 Loading Base Image Software with DVD (cont’d)
6.) Select one of the options for loading the system. Select choice [a] to load the complete disk.
Refer to Figure 8-4 on page 8-6
- To select [a], the complete disk will be loaded. This option is recommended for
applicationsoftware upgrade.
NOTE: NOTE: When to select [a] to load complete disk, please ensure that any patient data on the disk
has been backed up.
- To select [b], only the bootable C: partition is loaded. This option is intended for recovery of a
system that will not boot up. It is not recommended for application software upgrade because
during upgrade process, the data on the system would possibly be impacted.
- To select [c] to quit system upgrade process
7.) Press “Yes” or “No” to continue.
If you select [a], ALL existing software and data will be erased. If backup has not been
performed, all data like Patient Database, System Configuration and User Configurations
(Customer Presets) will be lost.
Figure 8-4 Selection for loading the system
While the software install procedure is designed to preserve data, you should select
choice [b] to format disk C only.
Figure 8-5 Confirmation on loading the system