General Ultrasound system requirements
Ultrasound SystemCommon Service Information 3-3
Direction 5444964-100 English Rev. 5
General Ultrasound system
requirements
Contents in this section
• ‘Ultrasound system environmental requirements’ on
page 3-3
• ‘Electrical requirements’ on page 3-5
• ‘EMI limitations’ on page 3-18
‘EMI prevention/abatement’ on page 3-19
‘Probes environmental requirements’ on page 3-20
‘Time and manpower requirements’ on page 3-20
Ultrasound system environmental requirements
If the Ultrasound system is very cold or hot
When unpacking the Ultrasound system, allow the temperature
of the Ultrasound system to stabilize before powering up. The
following table describes guidelines for reaching operational
temperatures from storage or transport temperatures.
CAUTION
If the Ultrasound system is very cold or hot, do not turn on its
power until it has had a chance to acclimate to its operating
environment.
Table 3-1: Ultrasound system acclimate time

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
Power Supply100-240VAC, 50/60Hz
Display15" LCD
Doppler ModesColor Doppler, Power Doppler
Imaging ModesB-Mode, M-Mode
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Vascular, Cardiac, OB/GYN, Small Parts, Musculoskeletal
ConnectivityDICOM, USB, Ethernet
Probe Connectors3
PortabilityMobile
Operating Temperature10°C to 40°C
Relative Humidity30%

Related product manuals