Electrical safety tests
Ultrasound SystemCommon Service Information 4-15
Direction 5444964-100 English Rev. 5
Uninterrupted Power Supply (UPS)
NOTE: For all instructions in the “Electrical safety tests” section in case
of using a UPS (uninterrupted power supply) the terms outlet,
wall outlet, AC wall outlet and power outlet refer to the AC power
outlet of the UPS. In case of further available AC (or DC) power
outlets at the same used UPS, these must remain unused i.e.
not connected to any other devices.
CAUTION
Please observe that some Uninterruptible Power Supplies
(UPS) may not be medical devices! If the UPS is not a medical
device, it has to be located outside of the patient environment
(according to IEC 60601-1 / UL 60601-1).
1. Patient environment
1.5 m 1.5 m
2.5 m

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
Power Supply100-240VAC, 50/60Hz
Display15" LCD
Doppler ModesColor Doppler, Power Doppler
Imaging ModesB-Mode, M-Mode
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Vascular, Cardiac, OB/GYN, Small Parts, Musculoskeletal
ConnectivityDICOM, USB, Ethernet
Probe Connectors3
PortabilityMobile
Operating Temperature10°C to 40°C
Relative Humidity30%

Related product manuals