2-14 MAC 1200 resting ECG analysis system Revision D
2012250-095
Functional Description: PCB Control CS_CI
Keypad Interface
The keypad interface contains the control register for 8 keypad columns and the
receiving register for 7 keypad rows, and the control signals for the status LEDs:
LED_LIN, LED_BATT_, LED_START, and LED_STOP.
Using the matrix of 8 x 7, up to 56 keypads can be detected. Identification of the key
pressed is as follows.
The controller activates a column, activation is via low-level, then the row-
register is read to identify the pressed key by a low-level. This procedure is
repeated with the next column, till all columns have been activated.
Printhead Control
The printhead controller takes on the complete control of the 216-mm thermal
printhead with a line width of 200 mm.
The output rate to the printhead is 1000/second. The resolution in the Y-direction is
8 dots/mm, in the X-time axis up to 40 dots/mm.
Thermal Printhead Dot Control
The MPC855T prepares a complete dot column and sends it to the FIFO. An FPGA
reads out a complete column of the FIFO and generates the digital control signals for
the printhead, which are shifted in series.
The duration of heating a dot column is defined through the pulse width of the PWM
Signal HEAT that is generated from the FPGA too.
In addition, the duration of heating a dot column is influenced by the thermal
printhead monitoring.

Table of Contents

Other manuals for GE MAC 1200 ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1200 ST and is the answer not in the manual?

GE MAC 1200 ST Specifications

General IconGeneral
BrandGE
ModelMAC 1200 ST
CategoryMedical Equipment
LanguageEnglish

Related product manuals