Revision D MAC 1200 resting ECG analysis system 5-5
2012250-095
Troubleshooting: Basic Information
MAC 1200 Only Prints on the Upper or Lower Section of the Printout
Is the paper feed flap is only engaged on one side?
MAC 1200 Prints, but only Baselines are Printed Out
Are the electrodes applied correctly?
Are electrode cables plugged into the patient trunk cable terminal-box
correctly?
Is the patient trunk cable defective (e.g., RL defective)?
Are there contact problems at the patient input connector of the MAC1200?
If not: The PCB control defective.