Revision D MAC 1200 resting ECG analysis system 7-5
2012250-095
Service and Maintenance: Technical Inspections/Preventative Maintenance
Device Test Result Check
Output “Test Results” as detailed in Chapter 3, “Test Results”.
Main parameters:
All memory stores free from error?
ASIC test O.K?
Sample rate 1000?
Printhead voltage >18 V, battery charge O.K?
Printout clearly legible and without any lapses or interference?
RS-232 Interface Test
Carry out the “Interface Test” item “(1) Transmitting and receiving”, as detailed in
Chapter 3, “V24 Tests”.