Page 10 Protimeter MMS Moisture Measurement System
Surface Temperature Sensor is connected to the instrument and held against 
the surface of interest. Temperature values can be displayed in °C or °F (see 
section 2.5 for instructions).
T
DIFF is a useful function when investigating condensation, as it tells the user 
how many degrees a surface is above or below the prevailing dew point 
temperature. When a surface exceeds the dew point by more than 3°C, T
DIFF is 
displayed with the message NO CONDENSATION. When a surface exceeds 
the dew point by 3°C or less, T
DIFF is displayed with the message AT RISK, NO 
CONDENSATION. When the surface temperature is equal to or lower than the 
dew point, a negative T
DIFF is displayed with the message CONDENSATION. 
The progress bar is activated when T
DIFF is less than 14°C above the dew 
point. The indicator moves from left to right to signify the increasing risk and 
degree of condensation. Table 3 below summarizes this functionality.
Table 3: Condensator Mode Readings
Temperature of Surface MMS Condensator Mode Display
>14°C above dew point T
DIFF, NO CONDENSTATION,
[inactive progress bar]
≤14°C but >3°C above dew point T
DIFF, NO CONDENSTATION,
[active progress bar]
≤3°C above dew point T
DIFF, AT RISK, NO CONDENSTATION,
[active progress bar]
≤ dew point T
DIFF, CONDENSTATION,
[active progress bar]