EasyManua.ls Logo

GE Protimeter MMS - Setup Mode - Selection and Use

GE Protimeter MMS
23 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protimeter MMS Moisture Measurement System Page 11
2.4.1 Sensor Installation
All humidity probes must be plugged into socket EC1 in the top (“nose”) of
the instrument. This includes:
POL4750 Hygrostick (30-100% RH)
POL7750-2 Humistick (1-100% RH)
The only accessories that should be plugged into socket EC2 on the side of the
instrument are:
BLD5804 Surface Temperature Probe
MMS-to-PC Data Logging Cable
Note: With older instruments, the Low RH Sensor (0-40% RH) BLD5812
was plugged into socket EC2 on the side of the instrument.
2.5 Setup Mode - Selection and Use
The Protimeter MMS instrument has a range of features that can be selected
by the user (see Table 4 on the next page).
To enter the Setup Mode, switch the instrument ON and disconnect any
attachments that may be connected. When the display shows three horizontal
dashes ( - - - ), press momentarily; the word SETUP then appears in the
bottom left corner of the LCD together with the code S 1 in the middle of the
display. Within three seconds, press
XX repeatedly to scroll through the setup
codes as detailed in Table 4 on the next page. Press
X to set the desired setup
option.

Related product manuals