Page 6 Protimeter MMS Moisture Measurement System
If this should occur, re-zero the instrument as follows:
1. Remove any attached accessories 
and select Moisture Search Mode 
(see section 2.1 for instructions). 
MOISTURE & REL is displayed.
2. With one hand, hold the instrument 
at least 12 in. (0.3 m) from all 
objects (see the figure to the right).
3. With your other hand, 
simultaneously press the 
XXand X 
buttons. Hold them down until the 
display shows nul, then Air. Quickly 
release the buttons and continue to 
hold the instrument as shown, well 
clear of your body and your other 
hand. Keep the instrument in free 
air (i.e. at least 12 in. (0.3 m) from 
all objects) while the display is 
showing Air.
4. After about 10 seconds, the display should return to normal and the 
instrument should read zero when held in free air.
You may now take normal Search Mode readings with the MMS.
2.1.1.1 Troubleshooting the Zeroing Function - The instrument 
changes mode when zeroing Is attempted
If the instrument changes mode when you attempt to zero it, you have most 
likely not pressed the two buttons at exactly the same time. Return to Search 
Mode and attempt to press and hold the two buttons simultaneously.
2.1.1.2 Troubleshooting the Zeroing Function - The instrument 
shows Air, but then displays Err or E30 and shuts OFF
If this situation occurs, repeat the zeroing procedure. If this does not solve the 
problem, it is likely that the temperature adjustment is outside the permitted 
range. If possible, allow the unit to return to normal room temperature (about 
20°C/68°F). If the unit will not re-zero at this temperature, the unit is likely 
faulty and requires repair/service.