Charge Batteries with the COMPIT TC4 Charger
Fig. 3-3 Red LEDs on charger
The batteries take up to 3 hours to recharge. Each of the
red LEDs corresponds to one of the charger
compartments (Fig. 3-3). During the charge cycle, the
corresponding red LED blinks at a slow rate
(approximately once per second). Note: If the red LED
does not light up, the battery may be inserted the wrong
way round. When the battery is charged, the LED is solid
red. The charging unit now trickle-charges the battery to
compensate for self-discharging.
The battery temperature is monitored in the charger.
When the temperature is too high, the LED is solid red
and the charger switches to trickle-charging.
If the battery is correctly inserted and the red LED does
not light up, the charger has identified a battery problem.
The charging current to the compartment concerned will
be cut off. Remove the battery and discard, observing the
applicable waste-disposal regulations.
Charge Batteries with the VARTA Charger
Fig. 3-4 Battery symbols and bars in the charger display
Insert 4 or 2 batteries. To charge only 2 batteries, insert
them in the two compartments on the right or on the left.
The batteries take up to 3 hours to recharge. Once the
batteries are inserted, battery symbols will appear in the
charger display where each symbol corresponds to one of
the charger compartments (Fig. 3-4). During the charge
cycle, the corresponding bar in the battery symbols
blinks. Note: If the battery symbols and bar do not light
up, only one battery may be inserted or the batteries are
inserted the wrong way round. When the batteries are
charged, the bars are permanently illuminated. The
charging unit now trickle-charges the battery to
compensate for self-discharging.
The battery temperature is monitored in the charger.
When the temperature is too high, the bar in the battery
symbol is permanently illuminated and the charger
switches to trickle-charging.
If the batteries are correctly inserted and the displayed
battery symbols show no bars, the charger has identified
a battery problem. The charging current to the
compartment concerned will be cut off. Remove the
battery and discard, observing the applicable waste-
disposal regulations.