Controls and Indicators
2001589-312 Revision B TONOPORT VI 11
2 Controls and Indicators
Fig. 2-1 Controls and indicators of TONOPORT VI
Functions of Button
1 Button : push to display the most recent pa-
rameter readings. The display will show:
- systolic value "S" (unit mmHg or kPa shown on the
display)
- diastolic value "D" (unit mmHg or kPa shown on
the display)
- pulse rate "HR" (unit min
-1
)
The same button is used
- to toggle between the day phase and the night
phase (section "Toggling Between Day and Night
Phase") and
- to program the BP monitor (chapter 3 "Setup")
2 Connection for blood pressure cuff
3 Calibration mark
4 Liquid crystal display (LCD)
5 Button : push to start and stop a measure-
ment and to confirm entries
6 (Rechargeable) batteries
7 Lid covering battery compartment
8 Nameplate
9 Port for connection to PC (RS232)
10 Port for connection to PC (USB)
1
2
3
4
5
6
7
8
9
10
Button
Mes-
sage on
display
Function
Push once H 1 clear memory
Push twice H 2 set date and time
Push 3 times H 3 select measurement protocol
Push 4 times H 4 activate calibration mode
Push 5 times H 5 display firmware version
Push 6 times H 6 select energy source
Push 7 times H 7 enable/disable audio signal
Push 8 times H 8 toggle pressure unit between
mmHg and kPa
Push 9 times H 9 select measurement method:
deflation measurement
method or inflation
measurement method
INFO
INFO
INFO
START
STOP