EasyManua.ls Logo

GE TONOPORT VI - Setting Time and Date; Selecting the Pressure Unit; Measurement Protocols; Enabling or Disabling the Audio Signal

GE TONOPORT VI
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Setup
18 TONOPORT VI 2001589-312 Revision B
Time and Date
Usually the BP monitors are set to the correct time and
date before delivery. Therefore, the time only needs to be
corrected to change between Standard Time and Daylight
Saving Time.
Setting Time and Date
Briefly switch TONOPORT VI off and on again and
wait for the time to be displayed.
Push twice: the display indicates "H 2".
Push : The year will be displayed, e.g. "2020".
If the indicated year is correct, confirm it with
or correct it with , then confirm with .
The month will be displayed, e.g. "03".
If the indicated month is correct, confirm it with
or correct it with , then confirm with
.
In the same manner, correct day, hour, and minute.
In the end, the time of day will be displayed again.
Selecting the Pressure Unit
Briefly switch TONOPORT VI off and on again and
wait for the time to be displayed.
Push 8 times: the display indicates "H 8".
Push : the display indicates "mmHg" or "kPa".
Either confirm with or switch to the other
option with , then confirm with .
Measurement Protocols
There is a choice of three different measurement
protocols:
Max. inflation pressure: day phase 250 mmHg
night phase 220 mmHg
Selecting a Measurement Protocol
Briefly switch TONOPORT VI off and on again and
wait for the time to be displayed.
Push 3 times: the display indicates "H 3".
Push : the display indicates "LLLL" (Selecting
a protocol automatically clears the memory. If you
want to retain the data, switch the BP monitor off.)
Push : the display indicates "P1" (protocol 1).
Either select program 2 or 3 by pushing or
confirm the selected protocol with .
Enabling or Disabling the Audio Signal
Briefly switch TONOPORT VI off and on again and
wait for the time to be displayed.
Push 7 times: the display indicates "H 7".
Push : the display indicates "0000" when the
audio signal is disabled, and "1111" when it is
enabled.
Either confirm with or switch to the other
option with , then confirm with .
INFO
START
STOP
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP
Protocol Day Phase
(7 a.m. to 10
p.m.)
Night Phase
(10 p.m. to 7 a.m.)
P1 every 15 minutes every 30 minutes
P2 every 20 minutes every 40 minutes
P3 every 30 minutes every 60 minutes
INFO
START
STOP
START
STOP
INFO
START
STOP
INFO
START
STOP
START
STOP
INFO
START
STOP

Related product manuals